Products
GDI Solutions provide the best technology and services in environmental and safety measurements.

Analytical Industries Inc - Premium GPR Series
Analytical Industries' high-accuracy portable oxygen analyzers are for applications which must be controlled at several defined measurement points.
The GPR-3500 model has a compact, rugged enclosure and includes stainless steel wetted parts, integrated flow meters and needle valves. It measures purity for inert gases, carbon-dioxide and oxygen.
The GPR-1200 compact portable trace oxygen analyzer features a 4-way valve that traps a gas sample in the sensor. This enables faster oxygen measurement times when moving from point to point in a process and also protects trace oxygen sensors from premature depletion due to exposure to ambient oxygen levels to increase their lifespan.
It all adds up to economical operation and low cost of ownership for this market-leading portable oxygen analyzer range.
Features and Benefits
- Rugged industrial enclosure
- Up to 30 days battery life
- 0-10 ppm ~ 0-100% O2 Measuring range
- Accuracy of better than 2% of range
- 24 to 32 months sensor life span (in normal use)
- Internal pump option
- 0-1V output
- XLT sensor options for CO2 background
Specification
GPR-1200 MS-2 | GPR-1200 | GPR-3500 MOVR | |
---|---|---|---|
Measurement range | 0-10, 0-100, 0-1000 ppm |
0-1, 0-10, 0-100, 0-1000 ppm,
0-1%, 0-25% |
0-100% |
Accuracy |
Less than 3% or 10 ppb under certain conditions
(Choose either 2% or less than 0.2 ppm (whichever is greater) under certain conditions) |
< 2% of FS under constant conditions | |
Response time(T90) | T90 reading in 10 seconds | T90 reading in 13 seconds | |
Sensitivity (LDL) | 5 ppb | 0.05 ppm | 0.1 % |
Linearity | < 1% of scale | ||
Sensor model | GPR-12-2000MS-2 | GPR-12-333 | GPR-11-120-OP |
XLT-12-333 CO2 gas mixture greater than 0.5% |
|||
Sensor life | 24 months in < 100 PPM O2 | 24 months in < 1000 PPM O2 | 24 months in < 100% O2 |
Calibration Interval | 30 days | ||
Inlet pressure | 0.34–2 barg (5-30 psig) with atmospheric vent | ||
Flow rate | 0.5-1.0 L/m (1-2 SCFH) | ||
Gas connections |
1/8" compression tube fittings
1/4” compression tube fitting on outlet (GPR-3500-MOVR) |
||
Wetted parts | Stainless steel | ||
Display | Graphical LCD 7 x 3.5cm (2.75 x 1.375”) | ||
Resolution | 0.001 ppm | 0.01 ppm | 0.1 % |
Enclosure | Painted aluminum NEMA 4X, 21.8 x 22.9 x 7.6cm (8.6 x 9 x 3") | ||
Weight | 5.4kg (12lbs) | ||
Compensation | Barometric pressure and temperature | ||
Signal output | 0-1 V | ||
LED indication | LOW BATT (72 hrs warning); charging mode | ||
Operating temperature |
5°C to 45°C (41°F to 113°F) (GPR sensor)
-10° to 45°C (14°F to 113°F) (XLT sensor) |
||
Power | Rechargeable battery (lead acid) | ||
Battery life | Up to 30 days on a single charge (40 hours with pump running) | ||
Area Classification | General purpose (safety zone) | General purpose only | |
ATEX: Ex II 2 G Ex ib